Assessment of Beam Quality in Some Radio Diagnostic Centers
dc.contributor.author | Kadiri, Sehad | |
dc.date.accessioned | 2025-07-17T09:01:46Z | |
dc.date.issued | 2025-07-17 | |
dc.description | Vlerësuan cilësinë e rrezeve X në qendra diagnostike, siguri pacienti. | |
dc.description.abstract | In general, ionizing radiation and in particular X-rays plays a very important role in medical examinations for diagnostic purposes. The rapid growth of the number of procedures for diagnostic reason in medicine as a result led to increase the number of equipment for X ray generation, which consequently has increased the patients dose and in the same time is increased the risk of malignant diseases. Measurement parameters of quality assessment for X radiation generators are done at nine different diagnostic centers of public sector in Kosovo during the period 2014-2015. Measurements consist on accuracy of tube voltage, reproducibility of output voltage, total filtration of the x-ray tube and delivered radiation dose. Measurements were performed using Multimeter PTW NOMEX and DAP (Dose Area Product) Meter PTW Diamentor CM. | |
dc.identifier.citation | Kadiri, S., Hodolli, G., Ahmetaj, S., & Bekteshi, S. (2016). Assessment of beam quality in some radio diagnostic centers. Journal of Chemical, Biological and Physical Sciences, 6(4), 1217–1222. | |
dc.identifier.uri | https://repository.aab-edu.net/handle/123456789/560 | |
dc.language.iso | en | |
dc.publisher | AAB College | |
dc.subject | Ionizing radiation | |
dc.subject | x-ray | |
dc.subject | quality assessment | |
dc.title | Assessment of Beam Quality in Some Radio Diagnostic Centers | |
dc.type | Article |